Daily maintenance, precautions, troubleshooting and handling of zirconia oxygen analyzer

Release time: 2023-02-27


  Daily maintenance, precautions, troubleshooting, and handling of zirconia oxygen analyzers

  1. Why can't calibration be performed immediately after putting the instrument into use?

  Answer: This is because: Within 24 hours of cold start-up, the indication is abnormal. After one day of use, calibrate with standard gas. This is because some adsorbed moisture or combustible substances may exist in the cold machine detector or newly installed detector. After the machine is heated, at high temperatures, this adsorbed moisture evaporates, and the combustible substances burn, consuming the reference air in the reference side battery, causing the oxygen content of the reference air to be lower than the normal value of 20.6%, resulting in a low detector signal, or even a negative signal, causing the measured oxygen content to be higher, or even greater than 20.6%. The measured value at this time is inaccurate. It should be waited until the moisture and combustible substances inside the detector are replaced with fresh air before accurate measurement can be made. Therefore, the zirconia detector needs to be heated for at least one day before calibration.

  2. Why is it necessary to regularly calibrate the analyzer?

  Answer: There are many interfering factors in the use of zirconia analyzers, such as aging of the zirconia tube, ash accumulation, and corrosion of the electrodes by SO2 and SO3. After running for a period of time, the performance of the instrument will gradually change, bringing errors to the measurement, so it is necessary to regularly calibrate the instrument! The calibration cycle is usually 1-3 months, depending on the use environment and usage of the instrument.

  During calibration, pure N2 cannot be used as the zero point gas. Usually, the zero point gas should be 10% of the full scale; the range gas is 90% of the full scale; dry air is used as the range gas on site; the zero point gas uses 100PPM O2. This is because below 100PPM zero point, the standard gas error has too much influence on the instrument, and the calibration purge time is too long and it is not easy to purge in place; the measured value uses the downward extension line of the measurement linearity. Practice has proved that our choice is clear and effective!

  3. Why should the instrument not be turned on and off easily?

  Answer: There are two reasons: First, because the zirconia tube is a ceramic tube, although it has a certain resistance to thermal shock, during the on-off process, due to rapid cooling and heating, large temperature changes may cause the zirconia tube to break. Therefore, as little unnecessary on-off operation as possible should be done; second, the thermal expansion coefficient of the platinum electrode coated on the zirconia tube is inconsistent with that of the zirconia tube. After being used for a period of time, it is easy to fall off during the on-off process, causing the impedance inside the probe to increase, or even damaging the detector. Shutdown should be cautious!

  4. Judgment of detector constant temperature

  Answer: Enter the menu and check whether the detector temperature and voltage are consistent. This helps to determine whether the heating and temperature control system is normal. When the detector temperature is much higher than the constant temperature, it indicates that the thermocouple is open-circuited. Because the converter has a broken couple protection circuit, once the thermocouple is open-circuited, it will generate a millivolt signal to replace the thermocouple signal, causing the detector temperature display to be higher, and causing the heating power supply to be disconnected to protect the detector from being burned. At this time, although the temperature is very high, the electric furnace is not actually heating. Measuring the resistance at both ends of the measuring thermocouple (the lead must be disconnected) can verify this. The normal resistance of the thermocouple should be less than 20 ohms.

  If the inspection finds that the temperature is lower than the constant value, this should consider that the heating is not performed, the heating wire is broken, or the temperature control system is malfunctioning or damaged.

  5. Measured value is too high

  Without considering the front-end factors, first consider whether the detector calibration port is leaking, and whether the flange screws are tightened; the instrument has not been calibrated for a long time or the calibration is incorrect.

  6. Measured value is too low

  The instrument needs calibration or needs to be recalibrated;

  Incomplete combustion of the boiler, combustible gases in the flue gas;

  7. Large fluctuation in measured value

  Detector aging, high internal resistance, poor electrode contact;

  High humidity or water droplets in the sample gas, vaporization in the detector;

  8. Extreme drift of measured value, signal exceeding range

  Detector components are damaged, such as zirconia tube breakage, open-circuit electrode leads, detector aging, temperature compensation resistor breakage (oxygen content 100%);

  9. Causes and symptoms of probe aging

  Generally, the probe aging we refer to refers to the aging of the zirconia detector, mainly manifested in the increase of internal resistance and the increase of background potential: (1) Increase of internal resistance

  In practical applications, the increase in internal resistance caused by probe aging is more common. Internal resistance refers to the input resistance between the two ends of the signal line. It is the sum of the lead resistance, the interface resistance between the electrode and the zirconia, and the volume resistance of the zirconia. Therefore, electrode volatilization, electrode shedding, and the reverse stabilization of the zirconia electrolyte (from stable zirconia to unstable zirconia) will all cause an increase in internal resistance. Measuring the internal resistance of the detector can determine its aging condition. According to experience, when the internal resistance increases to near its usage limit, a large signal jump phenomenon will occur, and some reactions are slow response phenomena. For these detectors, the background potential is not necessarily very large.

  (2) Increase in background potential

  Background potential is the additional potential of the battery. There are two factors that cause the increase in background potential: one is a permanent factor, which is parasitic on the battery, such as the corrosive effect of SO2 and SO3, and the asymmetry of the battery; the other is a temporary factor, such as electrode ash and poor air convection. Once the conditions improve, the background potential can be reduced.

  The increase in background potential often reflects the degree of detector aging. When the E0 value exceeds the maximum adjustment amount of the analyzer, it indicates that the detector is damaged.

  For example:

  A zirconia, the background potential at the time of leaving the factory is ±1mV, and its allowable change range is ±3mV. After half a year of use, the potential exceeds -3mV; after 18 months of use, it becomes: -29mV; this indicates that the detector has aged and needs to be replaced.

  It should be noted that the aging of some detectors is manifested in the increase of background potential, while some detectors, although aged, do not have this phenomenon, so we need to carefully analyze and treat them. When the cause of the increase in background potential is caused by temporary factors, as the use time increases, it is possible that the background potential will first increase and then decrease.

  The number of probes aging due to increased background potential is less than that due to increased internal resistance. A simple increase in background generally does not cause large signal fluctuations.

  10. Precautions:

  ① The sample gas needs to be pressure-controlled; typically, the inlet pressure to the instrument should not exceed 0.05MPa;

  ② When calibrating with standard gas, the standard gas flow rate should be controlled between 150-300ml/min;

  ③ All gas lines entering the instrument must undergo strict leak detection, and this work must be performed system-wide every six months when the instrument is in normal operation;

  ④ Before the gas path enters the instrument, it must pass through a physical filter, 10u; if gas blockage is found, the filter screen (filter) can be checked first;

  ⑤ Regularly clean the analyzer's fan filter, once a quarter; in harsh environments, frequent cleaning is required to prevent instrument overheating due to poor ventilation;

  ⑥ The instrument's installation location should be level and away from vibration sources, to prevent errors caused by uneven sample convection due to an unlevel detector;

  ⑦ The environment around the analyzer requires good ventilation; avoid enclosed spaces to prevent measurement errors caused by unbalanced oxygen levels;

  ⑧ Flammable gases are strictly forbidden around the analyzer, as this will seriously affect the detector's accurate measurement;

  ⑨ Since detection is performed at high temperatures, if the gas to be measured contains H2, CO, or CH4, these substances will react with oxygen, consuming some oxygen, reducing the oxygen concentration, and causing measurement errors. Therefore, this factor should be considered accordingly when the instrument measures gases containing flammable substances, to avoid inaccurate measurements.

  ⑩ When measuring corrosive gases, active carbon filtration should be used first.

  For other technical questions regarding the Zirconia Oxygen Analyzer, you can call for consultation: 18225808093

Recommended product

EXNFZRO®2 Explosion-Proof Zirconia Oxygen Analyzer Zirconia Analyzer


EXNFZRO®2 Explosion-proof Zirconia Oxygen Analyzer Significance of Measuring Oxygen Content with Zirconia Analyzer Flue gas oxygen content is one of the important monitoring parameters for boiler operation and an important basis for reflecting the completeness of combustion equipment and boiler operation. Its value is related to factors such as boiler structure, type and nature of fuel, boiler load, operating air supply conditions, and equipment sealing conditions. The smaller the oxygen content, i.e., the smaller the excess air coefficient, the more it indicates increased heat loss due to incomplete chemical combustion and mechanical incomplete combustion; the larger the oxygen content, i.e., the larger the excess air coefficient, the more it indicates excessive air supply. Excess air causes a drop in furnace temperature, which not only affects combustion but also carries away a large amount of heat and dust, increasing the calculated concentration of pollutant emissions. At the same time, a large air volume also increases the power consumption of flue gas exhaust. Controlling flue gas oxygen content is very important for controlling the combustion process and achieving safe, efficient, and low-pollution emissions. Application Scope The zirconia oxygen analyzer is an instrument used to monitor and control the oxygen concentration in combustion gases, boilers, and industrial furnaces. It is widely used in coal-fired power plants, gas boilers, waste incineration boilers, hazardous waste disposal incinerators, petrochemical cracking furnaces, metallurgical industry blast furnace hot blast stoves, oxygen-enriched combustion air pipelines, sintering furnaces, heating furnaces, heat treatment furnaces, metal material annealing furnaces, air separation oxygen production, food industry, cement kilns in the building materials industry, glass kilns, inert gas protection furnaces in the semiconductor industry, electrode material production furnaces in the lithium battery industry, and waste lithium battery recycling. It helps save energy and can also reduce CO2, SOX, and NOx emissions by controlling complete combustion, contributing to environmental protection, preventing global warming, and reducing air pollution. Anhui Tianfen Instrument Co., Ltd. is a manufacturer of zirconia analyzers, providing products for steel mills, chemical plants, power plants, waste power plants, and more. Their product range includes TKFX-ZOA zirconia analyzers, ZO-2000 desktop zirconia analyzers, MKFB-400-A/H explosion-proof zirconia analyzers, ZOA-300B zirconia analyzers, ZO-FIIB explosion-proof zirconia oxygen analyzers, CY-ZC, CY-ZDA zirconia analyzers, OXT3000 series zirconia oxygen content analyzers, integrated oxygen analyzers, corrosion-resistant zirconia analyzers, YB-88G zirconia oxygen analyzers, OXT1000 series zirconia analyzers, TKZO-5 zirconia analyzers, NFZRO2-5 zirconia analyzers, ZR402G zirconia oxygen analyzers, CE-2C, CE-2D zirconia analyzers, ZFK8R zirconia analyzers, hot blast stove zirconia analyzers for steel mills, special zirconia analyzers for hazardous waste, MK-400-A-FHC corrosion-resistant zirconia analyzers, abrasion-resistant zirconia analyzers, EXFZZRO2-88 explosion-proof zirconia analyzers, FZZRO2-5 zirconia oxygen analyzers, NFZO-05 high-temperature extraction zirconia analyzers, YB-88GJ-ZW zirconia analyzers, abrasion-resistant zirconia analyzers, oxygen analyzers, trace oxygen analyzers, steel mill zirconia analyzers, waterproof zirconia analyzers. Oxygen analyzers include the OAM-800 oxygen analyzer (zirconia) and the OAM-802 explosion-proof oxygen analyzer (zirconia), as well as the BOA-807 series (integrated) and BOA-807-R (split) series specifically for waste power plant boilers. Oxygen analyzers measure the oxygen concentration in flue gas and other non-combustible gases during combustion. Tianfen oxygen analyzers include the OAM-800 oxygen analyzer (zirconia) and the OAM-802 explosion-proof oxygen analyzer (zirconia). The standard oxygen analyzers include the integrated OAM-800 and the split OAM-800-R. They use a direct insertion online measurement method, can display measurement data and fault information on-site, and feature high accuracy, fast measurement, and low maintenance. The OAM-800 series features zirconia sensors with nano-coatings and various abrasion-resistant protective tubes. The explosion-proof oxygen analyzers include the integrated OAM-802 and the split OAM-802-R. These are PCEC-certified explosion-proof products suitable for boilers, heating furnaces, incinerators, converters, and other explosion-proof environments.
Learn More