Daily maintenance, precautions, troubleshooting and handling of zirconia oxygen analyzer

Release time: 2023-02-27


  Daily maintenance, precautions, troubleshooting, and handling of zirconia oxygen analyzers

  1. Why can't calibration be performed immediately after putting the instrument into use?

  Answer: This is because: Within 24 hours of cold start-up, the indication is abnormal. After one day of use, calibrate with standard gas. This is because some adsorbed moisture or combustible substances may exist in the cold machine detector or newly installed detector. After the machine is heated, at high temperatures, this adsorbed moisture evaporates, and the combustible substances burn, consuming the reference air in the reference side battery, causing the oxygen content of the reference air to be lower than the normal value of 20.6%, resulting in a low detector signal, or even a negative signal, causing the measured oxygen content to be higher, or even greater than 20.6%. The measured value at this time is inaccurate. It should be waited until the moisture and combustible substances inside the detector are replaced with fresh air before accurate measurement can be made. Therefore, the zirconia detector needs to be heated for at least one day before calibration.

  2. Why is it necessary to regularly calibrate the analyzer?

  Answer: There are many interfering factors in the use of zirconia analyzers, such as aging of the zirconia tube, ash accumulation, and corrosion of the electrodes by SO2 and SO3. After running for a period of time, the performance of the instrument will gradually change, bringing errors to the measurement, so it is necessary to regularly calibrate the instrument! The calibration cycle is usually 1-3 months, depending on the use environment and usage of the instrument.

  During calibration, pure N2 cannot be used as the zero point gas. Usually, the zero point gas should be 10% of the full scale; the range gas is 90% of the full scale; dry air is used as the range gas on site; the zero point gas uses 100PPM O2. This is because below 100PPM zero point, the standard gas error has too much influence on the instrument, and the calibration purge time is too long and it is not easy to purge in place; the measured value uses the downward extension line of the measurement linearity. Practice has proved that our choice is clear and effective!

  3. Why should the instrument not be turned on and off easily?

  Answer: There are two reasons: First, because the zirconia tube is a ceramic tube, although it has a certain resistance to thermal shock, during the on-off process, due to rapid cooling and heating, large temperature changes may cause the zirconia tube to break. Therefore, as little unnecessary on-off operation as possible should be done; second, the thermal expansion coefficient of the platinum electrode coated on the zirconia tube is inconsistent with that of the zirconia tube. After being used for a period of time, it is easy to fall off during the on-off process, causing the impedance inside the probe to increase, or even damaging the detector. Shutdown should be cautious!

  4. Judgment of detector constant temperature

  Answer: Enter the menu and check whether the detector temperature and voltage are consistent. This helps to determine whether the heating and temperature control system is normal. When the detector temperature is much higher than the constant temperature, it indicates that the thermocouple is open-circuited. Because the converter has a broken couple protection circuit, once the thermocouple is open-circuited, it will generate a millivolt signal to replace the thermocouple signal, causing the detector temperature display to be higher, and causing the heating power supply to be disconnected to protect the detector from being burned. At this time, although the temperature is very high, the electric furnace is not actually heating. Measuring the resistance at both ends of the measuring thermocouple (the lead must be disconnected) can verify this. The normal resistance of the thermocouple should be less than 20 ohms.

  If the inspection finds that the temperature is lower than the constant value, this should consider that the heating is not performed, the heating wire is broken, or the temperature control system is malfunctioning or damaged.

  5. Measured value is too high

  Without considering the front-end factors, first consider whether the detector calibration port is leaking, and whether the flange screws are tightened; the instrument has not been calibrated for a long time or the calibration is incorrect.

  6. Measured value is too low

  The instrument needs calibration or needs to be recalibrated;

  Incomplete combustion of the boiler, combustible gases in the flue gas;

  7. Large fluctuation in measured value

  Detector aging, high internal resistance, poor electrode contact;

  High humidity or water droplets in the sample gas, vaporization in the detector;

  8. Extreme drift of measured value, signal exceeding range

  Detector components are damaged, such as zirconia tube breakage, open-circuit electrode leads, detector aging, temperature compensation resistor breakage (oxygen content 100%);

  9. Causes and symptoms of probe aging

  Generally, the probe aging we refer to refers to the aging of the zirconia detector, mainly manifested in the increase of internal resistance and the increase of background potential: (1) Increase of internal resistance

  In practical applications, the increase in internal resistance caused by probe aging is more common. Internal resistance refers to the input resistance between the two ends of the signal line. It is the sum of the lead resistance, the interface resistance between the electrode and the zirconia, and the volume resistance of the zirconia. Therefore, electrode volatilization, electrode shedding, and the reverse stabilization of the zirconia electrolyte (from stable zirconia to unstable zirconia) will all cause an increase in internal resistance. Measuring the internal resistance of the detector can determine its aging condition. According to experience, when the internal resistance increases to near its usage limit, a large signal jump phenomenon will occur, and some reactions are slow response phenomena. For these detectors, the background potential is not necessarily very large.

  (2) Increase in background potential

  Background potential is the additional potential of the battery. There are two factors that cause the increase in background potential: one is a permanent factor, which is parasitic on the battery, such as the corrosive effect of SO2 and SO3, and the asymmetry of the battery; the other is a temporary factor, such as electrode ash and poor air convection. Once the conditions improve, the background potential can be reduced.

  The increase in background potential often reflects the degree of detector aging. When the E0 value exceeds the maximum adjustment amount of the analyzer, it indicates that the detector is damaged.

  For example:

  A zirconia, the background potential at the time of leaving the factory is ±1mV, and its allowable change range is ±3mV. After half a year of use, the potential exceeds -3mV; after 18 months of use, it becomes: -29mV; this indicates that the detector has aged and needs to be replaced.

  It should be noted that the aging of some detectors is manifested in the increase of background potential, while some detectors, although aged, do not have this phenomenon, so we need to carefully analyze and treat them. When the cause of the increase in background potential is caused by temporary factors, as the use time increases, it is possible that the background potential will first increase and then decrease.

  The number of probes aging due to increased background potential is less than that due to increased internal resistance. A simple increase in background generally does not cause large signal fluctuations.

  10. Precautions:

  ① The sample gas needs to be pressure-controlled; typically, the inlet pressure to the instrument should not exceed 0.05MPa;

  ② When calibrating with standard gas, the standard gas flow rate should be controlled between 150-300ml/min;

  ③ All gas lines entering the instrument must undergo strict leak detection, and this work must be performed system-wide every six months when the instrument is in normal operation;

  ④ Before the gas path enters the instrument, it must pass through a physical filter, 10u; if gas blockage is found, the filter screen (filter) can be checked first;

  ⑤ Regularly clean the analyzer's fan filter, once a quarter; in harsh environments, frequent cleaning is required to prevent instrument overheating due to poor ventilation;

  ⑥ The instrument's installation location should be level and away from vibration sources, to prevent errors caused by uneven sample convection due to an unlevel detector;

  ⑦ The environment around the analyzer requires good ventilation; avoid enclosed spaces to prevent measurement errors caused by unbalanced oxygen levels;

  ⑧ Flammable gases are strictly forbidden around the analyzer, as this will seriously affect the detector's accurate measurement;

  ⑨ Since detection is performed at high temperatures, if the gas to be measured contains H2, CO, or CH4, these substances will react with oxygen, consuming some oxygen, reducing the oxygen concentration, and causing measurement errors. Therefore, this factor should be considered accordingly when the instrument measures gases containing flammable substances, to avoid inaccurate measurements.

  ⑩ When measuring corrosive gases, active carbon filtration should be used first.

  For other technical questions regarding the Zirconia Oxygen Analyzer, you can call for consultation: 18225808093

Recommended product

Zirconia oxygen analyzer, oxygen analyzer


The zirconia oxygen analyzer is a high-precision, online monitoring device developed based on the principles of high-temperature oxygen ion conduction in zirconia ceramics and the concentration‑difference electromotive force. It serves as a core smart instrument for measuring oxygen content in industrial flue gases, optimizing combustion conditions, and managing environmental emissions. The device can directly measure gas oxygen concentrations in various furnaces and pipelines, offering real-time monitoring, stable and durable performance, and adaptability to harsh operating conditions. Widely applicable across multiple industries for production and environmental‑related operations, it is a critical tool for achieving energy savings, safe production, and compliance with emission standards. I. Company Profile Anhui Tianfen Instrument Co., Ltd. is a high‑tech enterprise specializing in the R&D, manufacturing, sales, and technical services of industrial process analytical instruments. With years of expertise in oxygen analysis, environmental monitoring, and industrial measurement and control, the company focuses on iterative upgrades of zirconia oxygen analyzers, gas analyzers, and industrial control equipment. Backed by mature production processes, rigorous quality‑control systems, and a professional R&D team, it provides customized monitoring solutions tailored to diverse industry requirements. Its products—known for precision, stability, durability, low power consumption, and ease of maintenance—serve a wide range of sectors including power generation, chemical processing, metallurgy, building materials, and environmental protection, earning high recognition from both the market and customers. Committed to quality and driven by technology, the company continuously supports industrial enterprises in achieving intelligent manufacturing, energy efficiency, and regulatory compliance. II. Core Technical Parameters This series of analyzers features standardized industrial‑grade specifications, meeting the detection needs of most industrial applications. Key performance indicators are outstanding and highly stable: the standard measurement range is 0–25% O₂, with custom ranges available upon request; basic system measurement error is ≤±0.5% FS, with high‑accuracy models reaching ±0.1% O₂; repeatability is ≤0.5% FS, placing its accuracy at an industry‑leading level; T90 response time is ≤5 seconds, enabling rapid capture of dynamic oxygen‑content changes; temperature control is maintained at 700°C ±0.1°C, ensuring stable operation of the sensing element; the device operates over a broad temperature range, tolerating ambient conditions from −20°C to 85°C, while high‑temperature probes can withstand flue gas temperatures up to 1,400°C. Signal outputs include standard 4–20 mA analog signals and RS‑485 digital communication compliant with HART protocol, ensuring compatibility with mainstream industrial control systems. Zero drift is limited to ≤±0.5% FS per 7 days, guaranteeing long‑term operational stability and significantly reducing failure rates. III. Key Technological Features 1. In‑situ direct measurement with ultra‑fast response: No sample preparation or pre‑treatment is required; the device can be inserted directly into the process pipeline for on‑site measurement, eliminating delays, blockages, and leaks associated with sampling lines. Its sub‑second response time provides real‑time feedback on combustion conditions, supplying precise data for system control. 2. High‑temperature and corrosion resistance, suitable for demanding environments: Featuring a highly dense, stable zirconia ceramic sensing core paired with a corrosion‑resistant, wear‑proof structural design, this analyzer withstands high temperatures, dusty conditions, and mildly corrosive flue gases, resisting erosion and aging while adapting to complex, harsh industrial settings. 3. Intelligent calibration and robust stability: Equipped with automatic zeroing and purging functions, the device exhibits minimal drift over extended operation, ensuring consistent and reliable data. 4. Easy installation and low maintenance costs: Available in modular, plug‑in configurations, it simplifies installation without requiring extensive modifications. With no consumable parts and infrequent calibration needs, it significantly reduces ongoing labor and replacement expenses. 5. Broad compatibility and strong adaptability: Standard industrial signal outputs enable seamless integration with PLCs, DCSs, and other industrial control systems, supporting remote data transmission and centralized monitoring, thus meeting the demands of smart production line upgrades. IV. Addressing Industry Pain Points 1. Resolving traditional detection delays and distortions: Conventional sampling‑based oxygen analyzers suffer from slow response times, clogged tubing, and condensation interference, failing to reflect real‑time furnace conditions. By contrast, this device offers in‑situ direct measurement with no transmission lag, delivering accurate and reliable data. 2. Overcoming challenges in high‑temperature, dusty environments: Many precision analyzers cannot endure the extreme heat, heavy dust, and high‑velocity flows typical of industrial furnaces, often resulting in sensor damage and data loss. This specialized device incorporates a high‑temperature, dust‑resistant structure, ensuring stable long‑term operation even under severe production conditions. 3. Tackling high energy consumption and incomplete combustion: Industrial furnaces frequently experience imbalances in air‑fuel ratios and inefficient combustion, leading to fuel waste, reduced productivity, and increased emissions. By precisely monitoring oxygen levels, this analyzer helps optimize air‑fuel ratios, improve combustion efficiency, and lower energy use and carbon footprints. 4. Alleviating burdensome and costly maintenance: Traditional instruments require frequent disassembly for calibration, filter replacements, and pipeline cleaning, imposing significant labor and expense. This device minimizes maintenance needs and lowers failure rates, effectively reducing overall production and operational costs. 5. Mitigating risks of non‑compliant environmental monitoring: Oxygen content in industrial flue gases is a key parameter for calculating environmental emissions. Manual measurements often suffer from delays and inaccuracies, increasing the risk of exceeding emission limits. Continuous, 24‑hour precise monitoring ensures compliance and controllability of emission data. V. Major Application Areas The device finds extensive use in various industrial combustion, flue‑gas monitoring, and atmosphere‑control scenarios, spanning several core industrial sectors: - Power generation: Online monitoring of oxygen levels in coal‑fired boilers and thermal power plant furnaces. - Chemical processing: Monitoring operating conditions of heating and incineration furnaces. - Metallurgy: Optimizing combustion in steel, coking, and heat‑treatment furnaces. - Building materials: Detecting flue‑gas composition in cement, glass, and ceramic kilns. - Environmental protection: Supporting oxygen‑level monitoring for industrial waste incineration and desulfurization/denitrification processes. Additionally, it is suitable for energy‑efficiency optimization and environmental monitoring in light‑industry, textile, food, and district‑heating facilities, and can also be employed for precise oxygen‑concentration control in nitrogen‑protection and inert‑atmosphere applications. VI. Trademark Ownership Statement We hereby solemnly declare that the seven trademarks—ZIROX, EXNFZRO, TKFXZOA, TFEX, TFYHG, TFZRO, and TFYB—are duly registered with the National Intellectual Property Administration of China by Anhui Tianfen Instrument Co., Ltd. The company is the sole legal registrant of these trademarks and holds full, exclusive trademark rights, protected under the Trademark Law of the People’s Republic of China, the Regulations for the Implementation of the Trademark Law, and other relevant laws and regulations. The official registration numbers for each trademark are as follows: ZIROX (No. 84554887), EXNFZRO (No. 82544696), TKFXZOA (No. 82536162), TFEX (No. 64377345), TFYHG (No. 79839887), TFZRO (No. 79839454), TFYB (No. 82528679). Without formal written authorization from Anhui Tianfen Instrument Co., Ltd., no entity, organization, or individual may, in any commercial context—including production, manufacturing, sales, marketing, promotional activities, online postings, or business collaborations—unauthorizedly use, reproduce, imitate, alter, or misappropriate these trademarks. Nor may anyone employ marks that closely resemble these trademarks and could cause market confusion. For all instances of trademark infringement or unfair competition, our company will collect and preserve evidence, pursue legal action through complaints, lawsuits, and accountability measures, and rigorously hold infringers civilly, administratively, and criminally liable, resolutely safeguarding our legitimate intellectual property and brand rights.
Learn More