Hot blast furnace zirconia oxygen analyzer
Keywords: R&D tuning fork densimeter, zirconia oxygen analyzer
Category:
Product Details
Oxygen Transmitter Parameter Requirements
Display Screen: LCD LCD display, Chinese menu. Inductive buttons, no need to open the cover, operate directly on the glass panel.
Display content: Oxygen concentration, zirconium tube temperature, zirconium tube voltage value, output current value, sensor internal resistance, self-diagnosis information, etc.
Operating Language: Chinese
Power Voltage: 220VAC ± 10%
Power Frequency: 50/60Hz
Heating Power: ≤ 150W
Measurement Range: 0.01%-25.00% (From 5.00%~25% customizable settings).
Output Signal: 4 ~ 20mA
Signal Interface: Supports RS485/MODBUS protocol
Bluetooth Port: With expandable Bluetooth function in the menu
Damping Adjustment: Damping coefficient can be set in the menu
Measurement Accuracy: ± of the range 1% or reading ± 0.1% O2
Constant Temperature: 700.0 ℃± 0.2 ℃ ( 500~900 ℃ customizable settings).
Protection Class: IP66
Calibration Method: Manual calibration / Fully automatic calibration, two methods selectable via menu (automatic calibration requires a matching automatic calibration unit box).
Data Hold: Output hold function available during calibration
Automatic Purge: Automatic purge function in the menu, regularly purges accumulated ash from the filter automatically.
Installation Method: Wall-mounted
Range Drift: Within 12 months, the drift amount does not exceed the adjusted range by 0.2%
Stability: ≤± 0.05% (Instrument continuous verification 4h )
Self-diagnosis and Alarm Function: Equipped with self-diagnosis function, including thermocouple reverse connection, thermocouple open circuit, heating abnormal, oxygen concentration alarm, temperature alarm.
Response Time: 0.5 seconds 90% response. (Timed from when gas enters the calibration gas inlet and analog signal begins to change)
Junction Box Material: Metal (steel is recommended / die-cast aluminum).
The instrument has stable temperature control capability and long service life. The zirconium oxide oxygen transmitter is composed of electronic components with a microcontroller as the core , including data processing , oxygen content, temperature, alarm, etc. display , heater control , current output, alarm output, etc. Measurement errors can be corrected at any time, and output, control methods, and alarm points can be changed.
Oxygen Transmitter Service Life: ≥ 5 years
Probe Technical Parameter Requirements
Flue Gas Temperature: ≤ 800 ℃
Flue Gas Pressure: -10 ~ + 50Kpa
Flue Gas Flow Rate: 0 ~ 50m/s
Ambient Temperature: -40 ℃ ~ + 80 ℃
Response Time: ≤ 0.5s 90%
Process Time ( T90 )≤ 3s
Zirconia Cell Resolution: < 1ppm
Insertion Depth: 1000mm (Can be customized according to customer requirements)
Material: Protection Tube 310S
Protection Tube Diameter : φ 51mm
Probe Installation Angle: Horizontal or vertically downward installation
Flange: JIS-5K-65-FF
Heating Time: 5 ~ 15min (Constant temperature 700.0 ℃).
Sensor Electrode: Platinum electrode.
Pressure Resistance: Normal use without leakage under flue gas pressure in any combustion control condition (sensor factory test pressure 400kpa zero leakage).
Sensor Zero Potential: Less than ± 0.5mv In air, at a constant temperature 700.0 ℃). (Long-term stable, no drift)
Sensor internal resistance: less than ± 20 ohms (in air, at a constant temperature 700.0 ℃).
Calibration gas: 5% and 20.6% O 2 ,background gas is nitrogen ( 1%~10% any standard gas between can be used for calibration, and the linearity within the full range is less than 0.01% )
Reference gas: Air, no external reference gas source required
Standard gas flow rate: 150 ~ 300 ml/min
Probe filter: Standard dust-proof precision filter
Service life: ≥ 4 years