Zirconia Oxygen Analyzer Technical Parameters
Release time: 2022-12-06
Technical parameters of zirconia oxygen analyzer. Technical parameters for ZrO2-5, ZOY, ZO-12B, AZ20, and ZR22G zirconia oxygen analyzers are as follows.
Display: LCD
Operating Language: Chinese (English optional)
Power Supply: 200-245V AC 50~60Hz
Power Consumption: ≤150W
Output Signal: 4~20mA
Measuring Range: Factory default 0.01%-25%. Optional 0.01ppm-20000ppm, 0.01%-100%
Communication: HART/485 (optional)
Alarm: 1-way switch output (optional)
Instrument Accuracy: ±0.75% of reading or 0.05% O2, whichever is greater; Detection limit is –0.05% O2
Repeatability: ±0.5% of full scale
Basic Error: ≤±0.5% (full scale)
Stability: ≤±0.5% (instrument continuous calibration for 4h)
Response Time: 90% response within 5 seconds. (Measured from when the gas enters the calibration gas inlet and the analog signal begins to change.)
Probe Material: 304, 316L, 310S, HC (optional)
Protective Tube Diameter: φ45mm (default), φ32mm, φ38mm, φ51mm
Flange: JIS-5K-65-FF (standard), can be specified by customer (DN50/DN65/DN80/DN100)
Heating Furnace Resistance: 55-65Ω (other resistance values can be customized.)
Protection Level: IP65
Calibration Method: Semi-automatic calibration, full-automatic calibration optional.
Ambient Temperature: -10℃~55℃
Sample Gas Temperature: ≤700℃
Flue Gas Velocity: 0-50m/s
Sample Gas Pressure Range: -30kpa~﹢5kpa (high-pressure solution can be customized.)
Probe Installation Angle: Horizontal, vertical downward.
Calibration Gas: 5% and 20.6% O2, background gas is nitrogen. Calibration gas flow rate: 150-300 ml/min
Self-diagnostic Information: Thermocouple reverse connection, thermocouple disconnection, heating abnormality, upper and lower temperature limit alarm, upper and lower oxygen limit alarm.